Regulations on the Calibration of the Observation Instruments of Exclusive Observation Stations
1.中華民國八十一年七月十八日經濟部經(81)中標字第 022143 號令、 交通部(81)交航發字第 8124 號令會銜訂定發布全文 16 條 2.中華民國九十二年七月二十二日交通部交航發字第 092B000056 號令、 經濟部經標字第 09204609280 號令會銜修正發布名稱及全文 15 條; 並自發布日施行 (原名稱:基本氣象儀器校驗辦法;新名稱:專用觀測站觀測儀器校驗 辦法) 中華民國一百十二年九月十三日行政院院臺規字第 1121032982 號公告 第 2 條、第 6 條、第 7 條第 2 項、第 8 條第 2 項所列屬「 交通部中央氣象局」之權責事項,自一百十二年九月十五日起改由「交 通部中央氣象署」管轄
資料來源:全國法規資料庫(ChOrder.json,版本 2026/7/24 上午 12:00:00)・政府資料開放授權
This Regulations is enacted in accordance with the terms in Paragraph 6 of Article 20, Meteorological Act.
1The Central Weather Bureau (hereinafter referred to as CWB)of the Ministry of Transportation and Communications
2(hereinafter referred to as MOTC), the laboratory accredited by the Chinese National Laboratory Accreditation, or the
3laboratory accredited by a cross-recognized foreign country(hereinafter either of the three laboratories aforementioned is
4referred to as the Calibration Unit), may accept the application of an exclusive observation station to calibrate its equipped
5observation instruments according to Paragraph 1 of Article 20, Meteorological Act.
1The standard for calibration of observation instruments shall trace to the standard of the National Laboratory of Weights and
2Measures Standard, Bureau of Standards, Metrology, and Inspection, Ministry of Economic Affairs, or the national standard of
3a cross-recognized foreign country; the calibration cycle of observation instruments shall follow the rules of the individual
4Calibration Unit.
5The preceding rules of calibration shall be enacted separately.
1An exclusive observation station shall follow a self-decided calibration cycle of each equipped observation instrument and
2apply for an instrument calibration from a Calibration Unit to prove it be qualified for use, and then can be used.
3The error of an observation instrument is within the tolerant error as listed in the attached “Criteria for accreditation of the
4observation instruments” shall be considered as qualified.(see attached tables)
5The difference between the displayed value of the observation instrument under calibration and the standard value of the
6standard device is named the error of that observation instrument; the tolerant errors settled for each type of observation
7instruments are named the maximum permissible errors.
1Should the observation instrument(s) of an exclusive observation station shows an error caused by natural disaster or other
2causes, an application for calibration from a Calibration Unit is needed.
1Should the CWB find that the observation record of a certain observation instrument of an exclusive observation station had
2successive or obvious errors, CWB may inform that station to stop using that instrument and order to send it for calibration.
1An exclusive observation station shall fill in an application form and turn in a calibration fee when applying for a calibration of
2its equipped observation instrument.
3When apply for a calibration from the CWB, the calibration fee shall be charged in accordance with the provisions enacted by
4the MOTC according to Paragraph 2 of Article 30, Meteorological Act.
5When apply for a calibration from a Calibration Unit other than the CWB, the calibration fee shall be charged in accordance
6with the provisions enacted by that unit.
1The calibration of observation instrument(s) shall be conducted at a Calibration Unit; whereas the Calibration Unit may send its
2faculties to its place to conduct an in-situ calibration upon the request and application of the exclusive observation station under
3at least one of the following conditions:
41.Too big to move;
52.Adhered to the ground, a building, or other mechanic equipment and hard to tear apart;
63.It’s liable to be damaged or lose its accuracy while moving or other considerations.
7For applying to the CWB for a calibration, the applicant shall not only turn in the application fee (stipulated fees) in accordance
8with Paragraph 2 of Article 30, but also pay for the errand costs (including transportation fees), overhead fees, and/or other
9necessary expenditures.
1The observation instrument sent for calibration shall bear marks of the brand of production or factory logo, serial number of
2production, engraved scale, and the vernier (if any); the attached technical document or brochures shall also be submitted for
3calibration’s reference.
4Auto-recording types of instrument for calibration shall be submitted with its recording or displaying device(s); power supply
5exclusively used for the instrument shall be sent together.
1Should the observation instrument sent for calibration is totally damaged, seriously contaminated, or not useable, the
2Calibration Unit might reject the application.
The Calibration Unit may tear apart or ask the applicant to tear apart the instrument for partial calibration, if necessary.
1If the instrument sent for calibration with two (2) or more scales and any one of them fails to meet the criteria, the whole
2instrument shall be considered as a failed one.
1The Calibration Unit shall issue a certificate to those instruments that passed the calibration, or a notice to those stations with
2instruments failed to pass the calibration.
1The format of the application form, certificate, and notice aforementioned in this Regulations shall be determined by the
2Calibration Unit.
1This Regulations shall become effective on the date of promulgation.
2Criteria for accreditation of the observation instruments
3A. Atmospheric Pressure Measuring Instrument
4Unit : hectopascal(Symbol : hPa)
5Types of InstrumentMeasuring Range Maximum Permissible Error
6Precision Mercurial
7Barometer866~1050hPa
8(650~787mmHg)±0.33hPa
9(±0.25mmHg)
10Fortin-type Mercurial
11Barometer866~1050hPa
13(±0.5mmHg)
14Aneroid Barometer866~1050hPa
15(650~787mmHg)±0.67hPa
16(±0.5mmHg)
17Aneroid Barograph866~1050hPa
18(650~787mmHg)±1.33hPa
19(±1.0mmHg)
20Electronic Barometer866~1050hPa
21(650~787mmHg)±0.67hPa
22(±0.5mmHg)
23B. Temperature Measuring Instrument
24Unit : degree Celsius(Symbol : ℃)
25Types of InstrumentMinimal Scale/Name Measuring Range(℃) Maximum Permissible Error (℃)
26Glass-tube
27Thermometer
28nterval
29≦1/5℃-50~-20 ± 0.5
30痋?0 ~100 ± 0.3
31nterval
32≧1/2℃-20 ~+40 ± 0.3
33?0~100 ± 0.5
34hermometer
35in Iron Tube≦-20 ± 0.5
36痋?0 ± 0.3
37aximum
38Thermometer<0 ± 0.5
39~ 50 ± 0.3
40inimum
41Thermometer<-40 ± 1.0
42?0 ~-20 ± 0.5
43痋?0 ± 0.3
44Metallic
45ThermometerThermograph -20 ~+40 ± 0.5
46hermometer ± 0.5
47Electronic
48ThermometerThermograph -20 ~+40 ± 0.5
49hermometer ± 0.5
50C. Humidity Measuring Instrument
51Unit:percentage (Symbol:% RH)
52Types of InstrumentMeasuring Range
53(% RH)Maximum Permissible
54Error (% RH)
55Hygrometer and/or
56Hygrograph0 ~ 100 ± 5
57D. Precipitation Measuring Instrument
58Unit:milli-meter (Symbol:mm)
59Types of InstrumentMaximum Permissible Error(%)
60Rain Gage± 0.3
61Tipping-bucket Rain GagePrecipitation ± 3
62E. Wind-direction Measuring Instrument
63Unit:degree
64Types of InstrumentMeasuring Range
65(degree)Maximum Permissible
66Error(degree)
67Anemometer
68or
69Anemometrograph
700 ~ 360
71± 5
72F. Wind-speed Measuring Instrument
73Unit:meter per second (Symbol:m/s)
74Types of InstrumentMaximum Permissible
75Error(%)
76Propeller-type
77Anemometer/Anemometrograph± 5
78Three/four cup-type
79Anemometer/Anemometrograph± 5
80G. Sunshine-duration Measuring Instrument
81Unit:hours (Symbol:h)
82a. Minimum threshold:120W/m2
83b. Maximum Permissible Error:± 0.1h
84H. Solar-radiation Measuring Instrument
85Unit:watts per square meter (Symbol:W/m2)
86Maximum Permissible Error:± 5%